It is constructed in compliance with IEC 60695-2-10:2013, IEC 60950-1, IEC 60335-1, IEC 60884-1, IEC 60238, IEC 60695-2-11, IEC 60695-2-12, IEC 60745-1, IEC 60400, IEC 60730-1, IEC 60598-1, IEC 61050, IEC 60320-1, IEC 61058-1, IEC 60742, IEC 60669-1, UL 746A, BS EN 60695-2-1, BS 6458, NF C20-921-1, IS:11000, EN, TS, SIST …
http://webstore.iec.ch/preview/info_iec60695-2-10%7Bed2.0%7Db.pdf
As it is usually the case with other flammability tests also with this test it is not easy to obtain repeatable results. Therefore we designed GLOW WIRE TEST APPARATUS WITH FUME CUPBOARD with the goal of improved repeatability of test results and ease of operation. The BOLD features below are some of the advantages that were introduced to reach these goals.
T4-11 GLOW WIRE TEST APPARATUS WITH FUME CUPBOARD – MOTOR DRIVEN AUTOMATIC is intended for testing the flammability and ignitability of components and materials. It includes electronic voltage stabilizer – current regulator, A‑meter, thermometer, motor drive, timer, and FUME CUPBOARD.
Apparatus consists of:
The Control Unit with the electric circuit is built into the housing of the bottom part of the fume cupboard, so that all controls handy and in not contaminated position. A mechanism for performing the test is also located below the baseplate and therefore protected from contamination. Measuring instruments and components for controlling the operation of the apparatus are located on the front plate of the apparatus. The source of heating current is a toroidal transformer that is controlled (its primary side) through a microprocessor-controlled electronic circuit that operates as a power supply stabilizer. Potentiometer, that is located on the front plate, is intended for heating current regulation (adjustment).
To enable more precise current (temperature) regulation ten‑turn potentiometer with 1000 digital dial is built‑in. When the requested temperature stabilizes, push‑button should be pressed and by that voltage stabilization becomes active.
The electric circuit also contains control and switching elements for controlling the electric motor drive that slowly moves the tested sample to the wire loop after its activation by a push-button on the front plate. Just before the sample contacts the glow wire loop it slows down, and at the time of contact, the timer is activated which, after 30 seconds (or other set time), returns the carriage with the sample to the start position.
Mechanism with a wire loop, thermocouple, and motor drive is located under the base plate of the fume cupboard. Carriage with sample mounting support is so designed that mounting support can be easily removed for the purpose of sample mounting. All regulating elements of the mechanism are equipped with handle‑screws to enable easy operation. Wire loop and thermocouple are easily exchangeable. Apparatus detects loop penetration depth and time at which it reached 7 mm. It also has a scale for flame height reading with zero-point adjustments.
PROTECTION
The apparatus is protected against overheating, short circuit and break off in the circuit. Three limit comparators are built In the thermometer, of which the first is used as protection against too high temperature that could shorten the lifetime of the thermocouple. It switches off the heating of glow wire at a temperature of 1000°C, which is 40°C over the max. temperature that is requested by standard (960°C).
In case of short circuit between the loop carrying rods or between the heating current leads in manual or stabilization mode, processor detects irregular conditions and turns the heating off. This is indicated by the blinking of signal push‑button. After the cause of interruption has been removed, heating can be turned on by pressing “STOP” and “START” push‑button. In case of break off in the circuit or bad connection in the heating current in manual or stabilization mode, the processor detects irregular conditions and turns the heating off. This is indicated by the blinking of signal push‑button. Heating can be turned on by pressing “STOP” and “START” push‑button.
Tabletop apparatus, solid case, Non-sensitive, scratch-resistant surfaces through powder-coating, side panels, Al extrusion Frame, adjustable case feet with an anti-slip protection, front plate anodized aluminum 2,5 mm.
All parts of the mechanism are nickel-plated or made of stainless steel (AISI 304, DIN W. No. 1.4301 X5CrNi18-10) http://en.wikipedia.org/wiki/Stainless_steel.
Voltage stabilizer ensures better repeatability of test results, to avoid the following possibility: during the setting of heating current needed for temperature (e.g. 960 deg. C), the network voltage was 230V. At that voltage-current 130 A results in 960 deg. C. Then you begin the test. The voltage in the network drops (or oscillates) to 225V, which is approx 2% and this would result in different current (e.g. 127A) and different temperatures (18 deg. C difference).
Of course, if you do not like or do not need stabilization you can switch it off.
Due to the high currents involved, the wire loop is hard soldered to two stainless steel discs with smooth and large contact surfaces that ensure constant contact resistance (which means constant current during the test). This is important because if it was not hard soldered, the contact resistance between loop and stud (column) can change because of oxides formation due to the heat. That would result in a change of current and temperature which are the main parameters of this test. Again repeatability of test results improves.
The temperature stability depends on air movements around the glow wire loop. Control through the elements on the front plate that is outside the fume cupboard enables the operation of the glow wire test apparatus inside a closed test chamber. The test chamber enables draught-free conditions, has a volume of more than 0,5 m3, and has black painted inner walls as per standard requests. Draught-free conditions again mean more stable temperature and better repeatability of test results.
After the conditions stabilize, you can press the START button on the front plate, the complete test is executed automatically, just record the height and duration of the flames and start the ventilator after the test is finished.
Power supply | 230(120)V ±10%, 50(60)Hz ±2Hz (Other optional) |
Temperature adjustment | 370°C-1000°C in two ranges |
Control and Stabilization system | microprocessor, Voltage stabilization- current stabilization |
Thermocouple | Ni‑Cr‑Ni Type K Ø 1.0mm -200 to 1000°C |
Thermometer | digital programmable, accuracy ±0.2%, -200 to 1000°C |
A‑meter | digital A‑meter 150 A RMS |
Heating transformer | toroidal‑separating 500VA, secondary winding 2.5 V, 150A |
Dimensions W x D x H: | 645x 1105 x 1135 mm, Weight: 99 kg |
1 pcs. T4-08F WIRE LOOP FOR GLOW WIRE TEST APPARATUS FOR THERMOCOUPLE DIA. 1 mm –pre annealed for 10 hours
1 pcs. T4-08E THERMOCOUPLE DIA. 1 mm FOR GLOW WIRE TEST APPARATUS
1 pcs. Timing device with resolution better than 0,2 s
20 pcs. Silver foil with a purity of 99,99 %, approximately 2 mm2 and 0,06 mm thick for calibration of the temperature measuring system, purity certificate included
1 pcs. Mirror polished stainless steel plate that enables easy cleaning of possible burned plastic remains
1 pcs. Flat smooth wooden board, having a minimum thickness of 10 mm
200 pcs. Paper – wrapping tissue, as specified in 6.86 of ISO 4046, soft and strong, lightweight wrapping tissue of grammage 28 g/m2
T4-08E THERMOCOUPLE DIA. 1 mm FOR GLOW WIRE TEST APPARATUS
T4-08F WIRE LOOP FOR GLOW WIRE TEST APPARATUS (for TC dia. 1 mm)
T4-08D SILVER FOIL 2 mm2 – 20 pcs.
Certificates for force, dimensions of wire loop and temperature are available on request but are not included in the price.
Warranty: 2 years